Operators Guide Inspector 4000
002-5564 RJS, Minneapolis, MN
10
Verifying that Unit is Calibrated
Selecting CALIB sets up the mode to measure reflectance. This also allows you to
check the calibration of the unit by placing the wand on a template (white or black
field) and verifying that the reflectance value matches the stated values.
Verify that reflectance values are within 5% for white and 2% for black. If the values
are not within these tolerances you need to calibrate the unit; otherwise, press Enter
and Select at the same time to navigate back to the Main Menu.
Calibrating the Unit
After selecting Calib from the main menu, the reflectance menu will display.
Inspector 4000
Reflectance XX%
From the reflectance menu press Enter, the Calibrate menu will display:
Calibrate
When the Calibrate menu is displayed, proceed with the following calibration steps:
Place Optics on
White/Enter Key
Place the wand on the white section of the calibration plaque with the shoe guide flat
against the surface. Press Enter and hold auto-optic scanner head still until you hear a
chirp sound.