Rjs Inspector D4000 Laser (FIRMWARE version A.05 and E Instrukcja Użytkownika Strona 76

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 81
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 75
Appendix K Operators Guide
RJS, Minneapolis, MN 002-5568
70
Reference Decode
Each symbology type specifies a specific decoding method to be used in determining
overall symbol grade.
This parameter grade can be A or F. (A is pass and F is fail.) If this parameter
is F the overall symbol grade will also be F regardless of any other parameter. If this
parameter is A the lowest of the other parameter grades determines the overall symbol
grade.
Reflectance
A measure of the amount of light reflected from an illuminated surface.
Refl(min) / Refl (max)
This parameter grade can be A or F. The reflectance value of the lightest space in
a symbol must be equal to or greater than twice the reflectance value of the darkest
bar.
Refl(max) = or > Refl(min) x 2 if yes A else F
Reflectance Calibration Standard
A standard or known reflectance value, usually printed on the back of a template or
plaque. Barcode readers and verifiers are calibrated for reflectance using these known
values.
Required PCS
This is the minimum PCS percent required for the symbol.
Resolution
The dimension of the smallest code element that can be printed; the higher the resolution
the clearer the image.
Przeglądanie stron 75
1 2 ... 71 72 73 74 75 76 77 78 79 80 81

Komentarze do niniejszej Instrukcji

Brak uwag